User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: FLUKE 8200A PERFORMANCE TESTS DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 15 NUMBER OF LINES: 53 CONFIGURATION: Fluke 5100B Fluke 8506A CONFIGURATION: ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {PRELIMINARY OPERATION} 1.002 DISP The performance tests should be conducted under the 1.002 DISP following conditions: 1.002 DISP Ambient Temperature 23 +/-5 degrees Celsius 1.002 DISP Relative Humidity <70% 1.003 DISP CONNECT THE 8200A TO THE AC LINE AND TURN IT ON 1.004 DISP Select Range/Function 1V dc with filter 1.005 DISP CONNECT A SHORTING JUMPER BETWEEN INPUT TERMINALS 1.006 EVAL {DOES THE DISPLAY READ -.0001 TO +.0001?} 2.001 DISP DISCONNECT SHORT AND CONNECT A TEST CABLE TO HI & LO 2.001 DISP INPUT TERMINALS OF 8200A 2.002 HEAD {DC VOLTAGE VERIFICATION} 2.003 ASK- R 2.004 5100 1 1V 0.01/ 0.01% E 2W 3.001 5100 1 -1V 0.01/ 0.01% E 2W 4.001 ASK+ R 4.002 5100 10 10V 0.01/ 0.01% E 2W 5.001 5100 10 -10V 0.01/ 0.01% E 2W 6.001 5100 100 10V 0.01/ 0.01% 2W 7.001 5100 100 -10V 0.01/ 0.01% 2W 8.001 5100 100 100V 0.01/ 0.01% 2W 9.001 5100 100 -100V 0.01/ 0.01% 2W 10.001 5205 1000 1000V .01/ .01% E 2W 11.001 5205 1000 -1000V .01/ .01% E 2W 12.001 5100 A 0V 0.0001U 2W 13.001 5205 A 1000V .02U E 2W 14.001 EVAL {DID THE 8200A AUTORANGE TO THE PREVIOUS TEST?} 15.001 HEAD {FILTER CHECK} 15.002 DISP SELECT UUT RANGE/FUNCTION TO 1 VDC, WITH FILTER 15.003 5100 6.3V 100H S 2W 15.004 EVAL IS THE READOUT {.0000 +/-.001 WITH THE FILTER ON?} END